6 Device PowerSupply General Purpose Power Supply (GPDPS)
160 System Reference, January 2001
The channels of one ganged group have to be in sequence
(e.g. DPS11, 12, 13, 14, 21 etc.).
Note that it is not possible to gang different types of power
supplies together.
For details on cable assembly on the DUT board refer to
“Routing DPS Lines” on page 163.
Setting up Ganged GPDPS Channels
Basically, ganged GPDPS channels can be set up in the Pin
Configuration window. Ganging GPDPS channels is done
by assigning the same pin ID to these channels. Combine a
consecutive number of “slave” channels to a “master”
channel. The numbering of the slave channels has to
follow sequentially from the number of the master
channel. The first channel in this consecutive row of
GPDPS channels is automatically set as the master
channel. For details, see the Test Setup manual.
On the firmware level, the commands DFPS and DDSL are
used in this conjunction.
DFPS <master_channel>,<polarity>,<(pin)>
DDSL <number_of_slave_channels>,<(pin)>
In the following example, slave channels 12, 13 and 14 are
ganged with master channel 11 for pin Vee:
DFPS 11,POS,(Vee)
DDSL 3,(Vee)
NOTE Multiply the value of the load capacitor with the number of
ganged channels.
Correctly Positioning Ganged GPDPS Connections
All ganged channels must be allocated sequential
numbers. For example, if you want to configure three
GPDPS channels for a supply pin, and the first channel is
DPS11, the following channels must have channel
numbers DPS12 and DPS13.